Multi 1970i

Operation

25

Note
At this point, the AutoCal TEC calibration can be terminated with <M>.
This corresponds to a single-point calibration. To do this, the instru-
ment uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the
asymmetry of the electrode.

5

When the measured value is stable, Ct2 appears.

6

Thoroughly rinse the electrode with distilled water.

7

If required, enter the temperature of the second buffer solution
with <

> <> while keeping the <RUN/ENTER> key de-

pressed.

8

Immerse the pH electrode into the second buffer solution.

9

Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:

O

mV

S

/

1

Sal

O

LoBat

nLF

Lin

Oxi

Tref25

Tref20

ARng

8

Time

Year

Ident

No.

Day.Month

Baud

Sal

cm

1/

K

/

%

RCL

REL2

pH

m

cm

S/

%

mg/ l

2 4 8

°

C

TP

REL1

AR

Cal

TEC

Auto

/

1

Sal

S

6 9

LoBat

nLF

Lin

Oxi

Tref25

Tref20

8

Time

Year

Ident

No.

Day.Month

Baud

Sal

cm

1/

K

/

%

RCL

REL2

pH

m

cm

S/

%

mg/ l

2 4 8

3

°

C

TP

REL1

mV

AR

Cal

TEC

Auto

ARng

pH

7 0 0

pH

Store

AR

1

LoBat

nLF

Lin

Oxi

Tref25

Tref20

ARng

8

Time

Year

Ident

No.

Day.Month

Baud

Sal

cm

1/

K

/

%

RCL

REL2

pH

/

mV

m

cm

S/

%

mg/ l

Sal

O

S

pH

C

2 4 8

T 2

°

C

TP

REL1

Cal

TEC

Auto