5 d. o. calibration, D. o. calibration – Xylem MULTI 1970I PORTABLE MULTIPARAMETER METERS User Manual

Page 34

Advertising
background image

Operation

Multi 1970i

32

4.4.5

D. O. calibration

Why calibrate?

D. O. probes age. This changes the slope of the D. O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.

When to calibrate?

z After connecting another D. O. sensor
z When the sensor symbol flashes (after the calibration interval has

expired).

Calibration procedure

The calibration is performed in water vapor-saturated air. Use the
OxiCal

®

-SL air calibration vessel (accessory) for the calibration.

AutoRead

The calibration procedure automatically activates the AutoRead func-
tion. The AR display indicator flashes. The calibration process is fin-
ished when AR stops flashing.

Displaying the calibra-

tion data

You can view the data of the last calibration on the display. The pro-
ceeding is described on page 50.

Printing the

calibration protocol

The calibration protocol contains the calibration data of the current cal-
ibration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 53).

Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.8.2
before calibrating. After a valid calibration, the record is printed.

Sample printout:

CALIBRATION PROTOCOL
02.03.02 14:19
Device No.: 12345678
CALIBRATION 02
Cal time: 02.03.01 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++

Advertising