Calibration schedule, Barometric pressure sensor, Pressure sensor – Hach-Lange ORBISPHERE 6110 Basic User Manual User Manual

Page 16: Temperature sensor, Carbon dioxide sensor

Advertising
background image

of the next calibration is also shown. A report is created on completion of any new calibration. The
last 10 reports can be viewed by selecting the Calibration reports option in the calibration screen.

Calibration schedule

The following table shows the recommended sensor calibration intervals based on an average of
500 package analyses per week. This proposed schedule can be modified according to operating
conditions.

Sensor

Interval

Sensor

Interval

Barometric pressure sensor

6 months

O

2

sensor

2 - 3 months

Flow meter

6 months

CO

2

sensor

6 months

Temperature sensor

6 months

Pressure sensor

6 months

Barometric pressure sensor

1. A calibration screen is displayed showing the current values measured by the instrument.
2. Using a precision certified barometer, measure the barometric pressure in the location where the

instrument is used and compare with the current barometric pressure value displayed. If the
values are the same press Cancel, otherwise enter the new value in the New barometric value
box and press Validation to store the new setting.

Pressure sensor

1. A calibration screen is displayed showing the current values measured by the instrument.
2. Using a precision certified barometer, measure the barometric pressure in the location where the

instrument is being used. Provided a barometric pressure sensor calibration has recently been
performed this measurement should be the same as the value displayed in the Reference
pressure
box. If this is the case press Enter, otherwise enter the measured barometric pressure
value in the Reference pressure box and press Enter.

3. Connect a precision certified pressure gauge (0-7 bar) to the gas out connection on the rear of

the instrument and measure the reference pressure. Enter this value in the Reference pressure
box and press Enter.

4. Press the Validation button at the bottom left of the screen to accept the new calibration and

complete the process.

Temperature sensor

1. A calibration screen is displayed and the instrument measures two temperature values and

checks the stability of the measurements. Provided the measurements are stable, you can
choose to accept or reject the new values. Press Validation to store the new setting.

Carbon dioxide sensor

A Verification option is also available for this sensor. The process is the same as a new calibration.
When verification is complete press the Ok button to exit. No report is produced for the verification
process. The CO

2

sensor is calibrated using pure CO

2

and air. CO

2

is supplied by the purge gas and

air is provided by an internal pump that takes air directly from the atmosphere.

1. A calibration screen is displayed showing the gas purity data of the main gas. If the value is

incorrect, enter the correct value in the Gas purity box.

2. When the measurement stabilizes press the Validation First Point button. A new measurement

is then initialized and the internal air pump automatically activated.

16 English