Positioning, Pre-installation, Physical installation – Hach-Lange ORBISPHERE 6110 Basic User Manual User Manual

Page 6: Post-installation

Advertising
background image

D A N G E R

Risk of finger trapping. A silicon insert is located on the transparent front door to assist in lifting and lowering the
door when adding or removing packages for analysis. Under no circumstances must this silicon insert be
removed.

C A U T I O N

The instrument is heavy (55 kg) so extreme care must be taken when handling, to avoid damaging the instrument
or inflicting personal injury. It is highly recommended to use a pulley system (or similar) attached to the lifting
ring located on the top of the instrument when unpacking or moving it.

Positioning

C A U T I O N

The instrument is heavy (55 kg) so extreme care must be taken with handling to avoid damaging the instrument or
inflicting personal injury.

Hach Lange recommends installing the instrument on a sturdy laboratory table that can more than
support the 55kg weight. Place it on a clean flat surface, and convenient to the power source. The
display screen should be at head-height for easy viewing and operation.

Pre-installation

1. You will need the antifoam cartridge filled with the antifoam liquid as part of the installation

process. Prepare this before starting the installation using the instructions in

Antifoam cartridge

preparation

on page 7.

2. Make sure you have suitable standards available before performing any calibration.
3. Your A1100 electrochemical sensor has been thoroughly cleaned and tested at the factory before

shipment and installed in the instrument. It has been shipped with a cartridge containing a
membrane and electrolyte pre-installed to protect the sensor head. The sensor must be removed
from the instrument and the cartridge removed and replaced with a new one prior to first use to
make it fully operational. These steps are described in the A1100 cleaning procedure for
6110 TPA
document supplied with the instrument. The new cartridge is included in the sensor
recharge kit, also supplied with the instrument.

Physical installation

An installation wizard will start automatically once power is connected to the instrument for the first
time.

1. Connect the instrument to a power source - see

Power supply

on page 7.

2. Turn the instrument ON. An instrument auto check routine will start automatically. As this is the

first time the instrument will have been powered up, the O2 sensor residual and the Antifoam
level
LEDs will always be red. This is normal as the system has not yet been fully commissioned
and these can be ignored at this stage.

3. The wizard will start automatically. Press the Installation button to proceed.
4. At the end of the process select the auto-check option. See

Instrument auto check

on page 9

for a more detailed explanation of the auto-check process and results.

Post-installation

1. The instrument has been delivered with English as the default language. However, a USB

memory stick is supplied with other available languages (German, Spanish, Chinese, and
Japanese). If you wish to keep English as the preferred language please proceed to the next step
(Step 2 below), otherwise follow these instructions for installing a new language:

• Make sure that Microsoft ActiveSync

®

is installed on your PC (a copy is available as a free

download from the Microsoft web site).

6 English