Basic user manual, Specifications, English – Hach-Lange ORBISPHERE 6110 Basic User Manual User Manual

Page 3

Advertising
background image

Specifications

Specifications are subject to change without notice

Specification

Details

Measurement range

O

2

1 ppb - 12 ppm

CO

2

0.75 - 5 V/V or 1.5 - 10 g/kg

Repeatability r

95

Total Package Oxygen

± 5 μg/L ± 10% whichever is the
greater

CO

2

on equilibrated packages at T

= 10 to 25°C (50 to 77°F)

± 0.05 V/V or 0.10 g/kg ± 2%
whichever is the greater

O

2

± 2 ppb ± 10% whichever is the
greater

Typical analysis time

About 4 minutes

Display units

O

2

concentration

ppb or ppm

CO

2

concentration

V/V, g/kg, g/L or %W

Pressure

bar, mbar, psia

Temperature

°C, °F or K

Operating limits

Package temperature

-2 to 30°C (28 to 86°F)

Package pressure

1.4 to 6.8 bar absolute (20 - 99 psia)

Ambient temperature

0 to 40°C (32 to 104°F)

Relative humidity

Up to 80%

Package setup

Maximum package height

340 mm (13.39 ins)

Minimum package height

90 mm (3.54 ins)

Minimum volume

150 ml

Material

Glass, PET or aluminum

Dimensions (L x W x H)

537 x 540 x 942 mm (21.1 x 21.3 x 37.1 ins)

Weight

55 kg (121 lbs)

Enclosure protection

IP20

Power requirements

100-240 VAC ±10% @ 50-60 Hz

Power consumption

Max. 250 VA

Purge gas

CO

2

with purity > 99.9% at 6 to 7 bar absolute (87 to 102 psia)

Purge gas consumption

0.4 mL/second (1.5 L/hour)

Forcing gas

Air or N

2

at 5.5 to 6 bar absolute (80 to 87 psia)

Forcing gas consumption

0.25 mL/minute in standby mode

European directives

Low voltage 2006/95/EC
EMC 2004/108/EC

Electromagnetic compatibility
standards

EN 61326: 2006

Safety standard

IEC/UL/CSA 61010-1

Laser product safety

IEC/UL/CSA 60825-1

English 3