• Using the USB cable supplied with the installation kit, connect your PC to the instrument via

the USB client connector. ActiveSync

®

will start automatically. By default ActiveSync

®

proposes to set up a partnership with the instrument. This is not required, so select NO before
continuing.

• Insert the supplied USB memory stick into your PC.
• Inside the Setup folder, double click on the setup.exe executable file. Follow the on-screen

instructions on your PC and then on the instrument to install the preferred language software.

2. Set the security levels, user ID’s and passwords as described in

Security and user management

on page 13.

3. Configure the instrument parameters as described in

Configuration

on page 10.

4. Perform a barometric sensor calibration as described in

Barometric pressure sensor

on page 16.

5. Perform an O

2

sensor calibration as described in

Oxygen sensor

on page 17. Wait until the

residual value of the sensor is low enough (controlled automatically by the instrument). This
operation may take around 1-2 hours depending on storage conditions.

Note: The other sensors are more stable and will not require calibration prior to use.

Power supply

W A R N I N G

Only use the supplied power cable. Removable power cables must not be replaced with inadequately
dimensioned power cables.

C A U T I O N

Check voltage requirement sticker on the rear panel of the instrument before connecting (100-240 VAC).

Connect the power cable supplied with the instrument to the socket at the rear of the
instrument (left in diagram) for AC power connection. To switch the instrument on and
off, press the rocker switch (right in the diagram) - I for On and O for Off.

Note: For optimum performance always keep the gas and power supplies to the instrument switched
on.

Emergency STOP button

The red STOP button is located on the right side of the instrument. Should the need arise to stop the
machine at any time during operation, press this button. An audible click can be heard as the button
locks in position. The instrument should then be disconnected from the power supply. Once
disconnected, unlock the button by turning it clockwise. The instrument can then be reconnected to
the power supply and restarted.

Antifoam cartridge preparation

An empty 280mL volume antifoam cartridge is delivered as part of the installation kit. A recharge
bottle of antifoam is part of the package along with the necessary tools to transfer the liquid to the
cartridge (100mL volume syringe, tubing and connectors).

Note: One antifoam recharge bottle should be enough to fill the cartridge at least 3 times and a cartridge
completely filled with antifoam is sufficient for approximately 3,000 measurements.

To optimize the performance and reliability of the instrument, it is strongly recommended to only use
the silicon antifoam supplied by Hach Lange (part number 33156). To fill the cartridge, proceed as
follows (numbers in the boxes correspond to the steps below):