Place the package to be analyzed on the circular platform, and set the tilt position as defined for this
package by turning the black knob on the front. Two red laser beams are activated as the front door
is raised. Where the two red beams intersect is where the package will be pierced.

For bottles, position the package so that the two red beams intersect in the centre of
the bottle top (illustrated above left).

For cans, position the package so that the two red beams intersect at position 1
(illustrated above right). This is the recommended piercing position. Alternatively,
position the package so that the two red beams intersect at position 2. If piercing at
position 2, turn the ring pull 90° left or right, in order to clear the piercing position. Do
not pierce the can with the ring pull in the position illustrated.

Make sure that the package is also hard up against the backstop. This may require realigning the
backstop by turning the locking wheel counter clockwise and adjusting its position to that defined for
this package. Lock it in place by turning the locking wheel clockwise finger tight.
Once the package is secure against the backstop and the piercing position aligned with the two red
laser beams, close the front door.
Start the measurement process by pressing the illuminated START button on the front right side of
the instrument base. A progress bar is displayed at the bottom of the screen showing the elapsed
time to date, and the estimated time for measurement completion. This bar is updated on an on-
going basis. No user intervention or action is required until the measurement process has completed.
However, you do have the option to Abort the process at any point.
Once complete, the measurement results are displayed. The three columns of data displayed are
those that have been defined in

Measurement output

on page 12 and are color coded (green

signifies the measurements are within the defined limits for this package, and red signifies outside
the defined limits).
It is possible to add 2 numerical values and 5 comments to these results by entering them in the text
boxes displayed down the left. The numerical values will be required if they have been defined as
part of a formula (see

Formula management

on page 12).

Press the Continue button to return to the main measurement screen, or press the Info button to
display more detailed results in a tabular format.

Analysis

Open data

This option allows you to view the measurement data from the last 1000 measurements. The data
elements displayed are those defined in

Table settings

on page 18. Use the Previous and Next

buttons to scroll through the data. Select one of the lines of data and the Info button will be made
available. Press the Info button to see more details of the measurement in a new screen.
Use the Exit button to return to the main analysis menu.

Table settings

This options allows you to define up to 10 data elements that the instrument will display for each
measurement. More data is recorded, but only 10 values can be displayed on screen. Define for
each of the 10 columns the data to view from the drop-down list of available data elements. Select
Validation to accept and save the selected data elements.