Advanced operations – Hach-Lange DR 3800 User Manual User Manual

Page 104

Advertising
background image

104

Advanced Operations

Second option:

1. Insert the blank cuvette/cell into the cuvette/sample cell holder.

Press Zero.

2. Insert the sample cuvette/cell into the cuvette/sample cell

holder. Press Read.

The newly plotted wavelength scan curves are shown in
black.

The absorbance or transmittance value and the associated
wavelength are highlighted in black.

3. Press Options and then Reference: Off in the Options menu

to select another scan to display on the same screen with the
current scan. Highlight the required scan number and press
Highlight Reference.

Note: After selecting a reference scan the Reference: Off key in the
Options menu turns into Reference: On.

Note: Only scans that have the same wavelength range and step can
be displayed using the overlay option. This process can be repeated
until all matching scans are displayed.

4. The reference curve is shown in orange. The absorbance or

transmittance value and the associated wavelength are
highlighted in orange.

Additionally the difference of the extinction and/or
transmission value between the two Scans (measured
Scan and reference Scan) is indicated/highlighted at each
position of the cursor.

Note: A black and an orange box are shown in the left corner of the
display. The orange box relates to the reference scan and the black
one relates to the current wavelength scan.

5. Press the black or orange small box in the left upper corner on

the screen to switch between the actual wavelength scan and
reference scan.