5 drift check, Advanced operations – Hach-Lange DR 3800 User Manual User Manual

Page 115

Advertising
background image

115

Advanced Operations

6.8.3.5 Drift check

The Drift Check test is used to test the stability of the instrument.

Note: The Drift Check runs 1 hour.

1. Press Drift Check in the Optical Checks menu.

2. Press

λ to input the wavelength.

3. Input the wavelength and press OK to confirm.

4. Press Start.

5. The Drift Check runs 1 hour.

This test takes a reading every minute for one hour. Every
fifteen minutes, linear regression is used to calculate the slope
(rate of change) for the previous fifteen minute interval.

Advertising