Installation

3.5 Cell compartments, cuvette/sample cell adapters, light shield and

Adapter Box

3.5.1 Cell compartments and adapters

The DR 3800 has two cell compartments (

Figure 3

). Only one

cuvette/sample cell type at a time can be used for a measurement.

Cell compartment #1

13-mm round cuvettes/cells

Note: Cell compartment #1 contains a barcode reader for
cuvettes/vials.

Cell compartment #2
Cell compartment #2 uses adapters to accommodate different
cuvette/sample cell types.

1-inch square or 50-mm rectangular cells (can be inserted
directly into the cell compartment without using an adapter).

Adapter A: 10-mm square cuvettes/cells

Adapter B: Pour-Thru and multi-path cells

Note: Pour-Thru Cell must be used with Adapter B, not Adapter C.

Adapter C: One-inch round cells and AccuVac

®

Ampuls

Note: One-inch round cuvettes/cells and AccuVac Ampuls must be used
with Adapter C, not Adapter B.