Advanced Operations

6. The last reading and slope for each 15 minute interval are

displayed. At the end of the hour, the overall values are
calculated and displayed. The test proceeds to completion.

7. Press Cancel to return to Optical Checks.

6.8.3.6 Verification kit

The Verification Kit (see

Section 9 on page 131

) is designed for

periodic monitoring of scattered light, photometric accuracy and the
wavelength accuracy of the spectrophotometers.

When results exceed allowable tolerances (given in the quality
control certificate to the test record), contact the manufacturer.

1. Press Verification Kit in the Optical Checks menu.

2. Press Nominal Values.

3. Press Edit.

An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the quality
control certificate, to the following specifications:

Stray Light

Photometrical accuracy

Wavelength accuracy

4. Press OK when all values are entered and the overview is

displayed.