Standard Operations
5.4.4.1 Take single wavelength measurements (single reading)
1. Insert the blank cuvette into the cuvette/sample cell holder.
Press Zero.
Note:The Read key is only active after the zero measurement has
been carried out.
2. Insert the sample cuvette into the cuvette/sample cell holder.
Press Read.
Note: For data storage, see
5.4.4.2 Take single wavelength measurements (continuous readings)
1. Insert the blank cuvette into the cuvette/sample cell holder.
Press Zero.
Note: In the reading mode ”Continuous“ only the Zero key is shown to
start the reading. The reading sequence is started automatically.
2. Insert the sample cuvette into the cuvette/sample cell holder.
3. Press Options and then the Store icon to store the displayed
data in the Data Log.
Note: For data storage, see