Standard Operations

5.5.2 Select the measuring range

Some tests can be used for different measuring ranges. After the
sample cuvette/vial has been inserted, a list of the different
measuring ranges is displayed.

Select the required measuring range by highlighting the appropriate
line.

Press Start Permanent if this measuring range is to apply to all
subsequent measurements.

Changing the standard setting
In the result display, press Options>More>Permanent: On. The
key will change to Permanent: Off.

5.5.3 Select the chemical evaluation form

The chemical form of the test result of a number of parameters can
be selected individually.

In the result display, press the unit (e.g. mg/L) or the chemical
representation of the evaluation form (e.g. PO

4

3–

–P). A list of

possible evaluation forms is displayed, from which the required
form can be selected. Press

OK

to confirm.

Another way of changing the standard setting is:

1. In the result display, press Options>More>Chemical Form.

A list of available evaluation forms appears.

2. Select the required chemical form and press

OK

to confirm.

Note: The selected chemical form is displayed, but does not become
the default. To change the default, see

section 5.5.3.1

.

5.5.3.1 Change of the default setting of the chemical form

1. Insert the blank or sample cuvette/cell (depending on the

working procedure) into the cell compartment.

2. In the result display, press Options>More>Chemical Form.

3. A list of available evaluation forms appear. Select the new

default setting.

4. Press Save as Default.

The current result and all further measurements will be displayed in
the new chemical form.