17

Installation

3.5.3 Use of the light shield for measurements

The light shield (

Figure 6

) prevents light interference when using

13 mm barcode tests and must be in place before measurements
can be taken in cell compartment #1. The light shield is required
only when using 13 mm barcode tests.

The DR 3800 is shipped with the light shield installed. Remove the
light shield before using cell compartment #2. The light shield can
be stored in the Adapter Box (

Figure 8 on page 18

)

.

Installation of the light shield

1. Open the cell compartment.

2. Insert the light shield so the arrow on the light shield points to

the left and the orientation tab fits the groove in the
compartment opening (

Figure 7

)

Figure 6 Light shield

Figure 7 Light shield installed

1

Light shield