Laurel Electronics LAUREATE SERIES DPM User Manual
Page 22
-20-
11.
DC VOLTS & AMPS
11.1
RANGE JUMPER SELECTIONS
Figure 11.1 DC Signal Conditioner
11.2
MENU SELECTION
Whenever the scale factor is 1.0 and offset is zero, the meter displays a direct readout of the
signal input in (milli)volts or (milli)amperes. In the following example, the meter is configured
for a full scale display of 0 to 20V or 0 to 20mA equals 0 to 20.000. Other ranges follow the
same setup format. Note that the decimal point selection does not affect the displayed value.
A full scale value of 20000 may be displayed as 20.000 milliamps or 20000 microamps. During
setup, it may be necessary to enable some menu items that are locked out. See Section 9 for
further information.
Voltage
Input Jumpers Required
200mV
E, b
2V
E, a
20V
F, g, b
200V
F, g, a
660V
F, h, a
Current
Input Jumpers Required
2mA
D, h, b
20mA
C, h, b
200mA
B, h, b
5A
A, h, b
a
b
A
B
C
D
E
F
g
h
Voltage
Input
E1
E2 E3
±200.00 mV
A
f
b
±2.0000 V
A
f
a
±20.000 V
B
h
b
±200.00 V
B
h
a
±600.00 V
B
g
a
Current
FS Input
E1
E2 E3
±2.0000 mA
A
e, h
b
±20.000 mA
A
d, h
b
±200.00 mA
A
c, h
b
±5.000 A
A a,b,h
b