Laurel Electronics LAUREATE SERIES DPM User Manual

Page 33

Advertising
background image

-31-

Press the

key to display

InPut (Input type selection).

Note: Selection of input type
& range must match jumper
selection in Section 16. 1.

Press

until AC U

(ac Volts) is displayed or

Press

to select 0.2V,

2.0V, 20.0V, 200.0V or
660.0V or

AC A (ac Amperes) is dis-
played.

2.0a, 20.0a, 200.0a (milli-
amps) or 5.0A (Amps)

Press to select value for
flashing digit.
Digit 1:
"0"= 20,000 cts. full scale
"2"=Same as "0" but LSD is
fixed zero.
"3"=2,000 cts. full scale
Digit 4:
Set to "0" for scale and offset

1 2 3 4 5

DIGIT SELECT KEY

VALUE SELECT KEY

Press the

key to display

SEtuP. (Basic setup). See
Section 9, Page 10 for de-
tailed description of selec-
tions for digits 1 through 5.

Press the

key to display

SCALE (Scale factor).

Press

to display value

and select left digit. Press
again to select another digit.

Press the

key to display

dEcPt (Decimal point).

Press to display decimal
point location.

Press

to select decimal

point location.

Use

to set digit values.

Set value and decimal to 1.0
or appropriate multiplier for
external shunts or CT's.

Press the

key to display

OFFSt (Zero offset).

Press

to display value

and select left digit. Press
again to select another digit.

Use

to set digit values.

Set value to 00.000. Deci-
mal point is fixed by Dec.Pt.

MENU KEY

Press

to display status

and select left digit. Press

again to select another

digit. Selected digit will flash.

Press the

key. Continue to press

(or

and

simultaneously) until rESEt

is displayed. The meter will now go to the operating mode and display the value of the input
signal.