Laurel Electronics LAUREATE SERIES DPM User Manual

Page 27

Advertising
background image

-25-

Press the

key to display

InPut (Input type selection).
Note: Selection of input type
& range must match jumper
selection in Section 13.1.

Press

to select thermo-

couple type J, K, T, E, R, S
and °C or °F scale (J°F, J°C,
K°F, K°C, t°F, t°C, E°F, E°C,
r°F, r°C, S°F, S°C)

DIGIT SELECT KEY

VALUE SELECT KEY

Press the

key to display

OFFSt (Zero offset).

Press

to display input

selected. Press

again

until tC (thermocouple) is
displayed.

Press

to display value

and select left digit. Press

again to select another

digit.

Use to set digit values and
set to 000.00 for °F and °C or

set value to 273.2 if °C is
selected to display in °Kelvin
and 459.7 if °F is selected to
display in ° Rankine.

Press the

key. Continue to press

(or

and

simultaneously) until rESEt

is displayed. The meter will now go to the operating mode and display the value of the input
signal.

Press the

key to display

SEtuP. (Basic setup)
See Section 9, Page 10 for
detailed description of selec-
tions for digits 1 through 5.

Press

to select value for

flashing digit.
Digit 1:
"0"= 0.1 degree resolution
"3"=1 degree resolution

1 2 3 4 5

Press

to display status

and select left digit. Press

again to select another

digit. Selected digit will flash.

MENU KEY

Advertising