Laurel Electronics LAUREATE SERIES DPM User Manual
Page 30
Advertising
![background image](/manuals/336109/30/background.png)
-28-
Press
until rAtio (Ratio-
metric operation) is dis-
played.
15
STRAIN GAUGES AND POTENTIOMETERS
DIGIT SELECT KEY
VALUE SELECT KEY
Press the
key to display
InPut. Note: Range selec-
tion must match jumper se-
lection in Section 15. 1 .
Press to select 0.2U, 2.0U
or 20.0U (200mV, 2V or
20Vdc).
MENU KEY
15. 2 MENU SELECTION
12.1
RANGE JUMPER SELECTIONS
Voltage
Input Jumpers Required
200mV
E, b
2V
E, a
20V
F, g, b
200V
F, g, a
660V
F, h, a
Current
Input Jumpers Required
2mA
D, h, b
20mA
C, h, b
200mA
B, h, b
5A
A, h, b
a
b
A
B
C
D
E
F
g
h
Voltage
Input
E1
E2 E3
±200.00 mV
A
f
b
±2.0000 V
A
f
a
±20.000 V
B
h
b
±200.00 V
B
h
a
±600.00 V
B
g
a
Current
FS Input
E1
E2 E3
±2.0000 mA
A
e, h
b
±20.000 mA
A
d, h
b
±200.00 mA
A
c, h
b
±5.000 A
A a,b,h
b