Laurel Electronics LAUREATE SERIES DPM User Manual
Page 24
![background image](/manuals/336109/24/background.png)
-22-
12.
PROCESS SIGNAL INPUTS
12.2
MENU SELECTION
Display in engineering units is most easily programmed using the coordinates of 2 points. There
are two methods. The first is to enter the 4 values (low signal input, desired reading at the low
signal input, high signal input, and desired reading at the high signal input) directly via the front
panel pushbuttons or the RS232 interface. The second method is to have the meter read the
signal input at a known low value and store that reading as the low in and read a known high
signal value and store that value as high in. The low and high known values are entered as the
lo rd and hi rd. An example of using the reading the input method of coordinates of 2 points
is shown for the load cell meter, Section 17. Selecting the reading method (menu item "config"
digit 2 set to 1) overides either method of scaling selected in "setup". The following example
is the 2 coordinate method of directly entering the 4 values.
12.1
RANGE JUMPER SELECTIONS
Voltage
Input Jumpers Required
200mV
E, b
2V
E, a
20V
F, g, b
200V
F, g, a
660V
F, h, a
Current
Input Jumpers Required
2mA
D, h, b
20mA
C, h, b
200mA
B, h, b
5A
A, h, b
a
b
A
B
C
D
E
F
g
h
Voltage
Input
E1
E2 E3
±200.00 mV
A
f
b
±2.0000 V
A
f
a
±20.000 V
B
h
b
±200.00 V
B
h
a
±600.00 V
B
g
a
Current
FS Input
E1
E2 E3
±2.0000 mA
A
e, h
b
±20.000 mA
A
d, h
b
±200.00 mA
A
c, h
b
±5.000 A
A a,b,h
b