Laurel Electronics LAUREATE SERIES DPM User Manual

Page 43

Advertising
background image

-41-

23.

DIGITAL INPUTS

Press the

key until

SEtup (Basic meter setup) is
displayed.

Press

to display status.

Press

again to select

digit. Selected digit will flash.
Digit 5: digital inputs A & B
at J1, Pins 5 and 4.

Press to select value
for flashing digit
0 A: Reset B: Meter Hold
1 A: Function Reset
B: Peak display
2 A: Meter Hold B: Peak
3 A: Meter Hold

B: Tare

4 A: Peak B: Tare
5 A: Tare

B: Reset

6 External Decimal Pts.1
7 External Decimal Pts.2

1 2 3 4 5

1 2 3 4 5

DIGIT SELECT KEY

VALUE SELECT KEY

MENU KEY

23.1

FUNCTION OF DIGITAL INPUTS

Tare Logical 0 - The present display value is set to zero and stored as an offset value.

Logical 1 - The displayed value is equal to the signal input minus the tare value.

Peak Display

Logical 0 - The peak value of the input signal is displayed.
Logical 1 - The present value of the input signal is displayed.

Hold

Logical 0 - The meter display and outputs are held at the last reading.
Logical 1 - The display and outputs are updated normally

Reset Logical 0 - The microcomputer reads and resets the meter to nonvolatile memory values

Logical 1 - The meter display and outputs operate normally.

Function Reset

Logical 0 - The microC resets peak to present value and resets alarms.
Logical 1 - The meter display and outputs operate normally.

External Decimal Points

Input A Input B

Decimal Pts 1

Decimal Pts 2

1

1

XXXXX

XXXX

.

X

0

1

XXXX

.

X

XXX

.

XX

1

0

XXX

.

XX

XX

.

XXX

0

0

XX

.

XXX

X

.

XXXX

23.3

MENU SELECTIONS