• Time, Test ends at the end of a preset time.
• Stable result, test ends if the results are stable within a value set by the

user.

• Time or stable result, combines both methods. Stop whichever occurs

first, a stable result or the end of a preset time.

For timed test the user sets a time of typically 5 minute to 99 minutes

Set stable results criteria. The test will stop when the weight changes less
than a preset amount in a preset time. The default setting is 0.002g in 15
seconds for PMB 53, and 0.02g in 15 seconds for PMB 202. The user can
set other values depending upon model.

Start Criteria

• Start Manually when the [Start] key is pressed
• Start Automatically when cover lowered.

Delays 5 seconds after cover is down before beginning.


Print settings:
To set if the analyzer should print and store results and when printing if the
compete test results or only the summary is printed and stored in the
analyzers internal memory.


Default settings (stored in PST 00, see section 6.3) are:
% Moisture
Single Temperature = 110ºC
Rapid = Off
Interval Time = 5 seconds
End Point Determination = Time and stable result

Time 60minutes, and

Stable PMB 53 = 0.002g/15 seconds or Stable PMB 202 = 0.02g/15 seconds
Start Criteria = manual start.
Print Test = on
Print Format = Complete

6.1.1 MANUALLY SETTING TEST PROCEDURES