7.0 RUNNING

A

TEST


If necessary the user can set a new set of values for a test setting or recall a
pre-stored set of values. See section 6. The last used set of values are
stored so they are always shown initially, even after power has been turned
off. That way if the last test settings are still valid the user can immediately
use them.

The basic operation after test settings are set is as follows:

Place an empty sample pan on the analyzer.
Press [Tare] to zero the weight.




Fill the sample pan with the sample. See Section 14 for assistance in how
to prepare a sample.

Place the filled pan on the analyzer.




Lower the lid of the

weighing chamber, if automatic

start has been selected the test will begin after a 5 second delay.

If Manual start has been selected press the [Start] key.

The test will begin.

If the rapid heat function has been set to MANUAL the display will ask if the
user wants the rapid heat function to operate during this test.

The display will show the last setting used, either:

MASS 23

ºC

0.000

g

MASS 23

ºC

5.670

g