After the test has finished the user can print either the complete test again or
only a summary by pressing the [Print] key before returning to normal
weighing.

The complete report is a copy of the data as shown above. If the summary
report is selected then the following is printed. Similar to the above except
without the interval readings.

MOISTURE TEST
FILE NAME: RES00032
DATE: 13/12/2008
TIME: 09:38:07
SER NO: XXXXXXXXXXX

USER NO: XXXXXXXXXXX

TEST NO: _ _ _ _ _ _ _ _ _

Preset: PST 05
Bread-4

Result: % Moisture
Heating: Step
Temp: Step 1 120 C
Time 1 04:00
Step 2 110 C
Time 2 90:00
Rapid Manual
Interval: 30 Sec
Stop: Time/Stable
60 Min
0.002 g
30 Sec
Start: Manual

INIT MASS: 12.341 g

*********** AUTO STOP ***********


LAST TEMP: 111C
TEST TIME: 05:30 Min
FINAL MASS: 11.820 g
MASS LOSS: 0.521 g
RESULT: 4.27 %M

Heading
File name used by analyzer to store result

From Analyzer
Add serial number

Add user number

Leave a blank line for users to fill in a number if
they wish, not from Analyzer memory
Settings of the test conditions

















Measured initial mass.

TIME STOP if timed out or MAN STOP if user
stopped

Summary of results



Extra LF’s at end of the report