At the end of 20 minutes the beeper will sound, read the temperature in the
chamber. Any time after the 20 minutes, press [Setup/Enter] to enter the
second temperature value.

Enter this value as before, Pressing [Setup/Enter] to confirm and store the
value.




The analyzer will return to normal weighing.

Leave the analyzer to cool

During the 20 minute soaking time you can force the timers to stop if you
observe the chamber temperature is stable by pressing the [Setup/Enter]
key. Then enter the temperature as described.

If the test is not completed as shown above, the temperature test will be
abandoned, the display will show “TEMP CAL FAIL” and the analyzer will
return to normal without the temperature calibration changing. The test will
be stopped if the second temperature is not entered with 30 minutes of
starting the high temperature test.

These values will be accepted only if they are within acceptable limits of the
values stored during the dealer calibration. See Section 7.2.2.2

SEC TEMP

000.0