Nova-Tech PMB Moisture Analyzer User Manual

Page 33

Advertising
background image

 

 

31

|

P a g e

© Adam Equipment Company 2009

To load this file press the [Setup/Enter] key again. Or to choose a different
file press the [Dspl] key to advance through the other files available, pressing
the [Setup/Enter] key when the desired file is shown.





After loading the file the display will show a Success message and return to
the USB screen.

To continue, the USB device should be removed. The analyzer will return to
normal weighing automatically. Or press the [Tare] key to escape back to
normal weighing.

To store or recall other test files plug the USB memory device back into the
analyzer.

The test files contain data to set the parameters of the test procedures in a
text file. All files will be in English regardless of the language selected for the
display. A typical test file will appear as:

Result=0;
Heat=1;
Ramp Temp=112;
Ramp Time=27:00;
Interval Time=05;
Stop mode=2;
Stop time=60;
Stable Range=0.002;
Stable Time=15;
Start mode=0;
Name=ABCD-0123
Rapid=1
savepos=01;

PST 06

PST