1 rs-232 hardware, 2 usb serial hardware, Serial interface – Nova-Tech PMB Moisture Analyzer User Manual

Page 58: Rs-232 hardware, Usb serial hardware

Advertising
background image

 

 

56

|

P a g e

© Adam Equipment Company 2009

11.0 SERIAL

INTERFACE

The analyzers have the ability to send or receive data over the serial
interface. Using either the RS-232 or the USB serial interface.

The weighing data can be sent over the interface either automatically or when
the user presses the [Print] key.

The user has control over what data is to be printed.

11.1 RS-232 HARDWARE

The RS-232 interface is a simple 3 wire connection. The input and output
connections are:

Connector:

9 pin D-sub miniature plug

Pin

2

Input to analyzer

RXD

Pin

3

Output from analyzer TXD

Pin

5

Signal ground

GND

Handshaking is not applied.

Baud rate:

2400, 4800, 9600, 19200, 38400

Parity:

NONE (=8N1), EVEN (=8E1) or ODD (=8 O 1)

All lines are terminated with carriage return and line feed (<CR><LF>).

In continuous output mode, the serial output format will be a single line in the
form “12.567 g<CR><LF>”.

11.2 USB SERIAL HARDWARE

The USB interface is a standard USB connection. To use the USB interface it
is necessary to plug the analyzer into the USB port of a PC.

The PC will recognise the new device and search for a suitable driver. If
none are found a message will be displayed to guide the user through the
installation of suitable software.