27

|

P a g e

© Adam Equipment Company 2009


6.2

STORE

TEST

SETTINGS


It will be possible to store up to 49 test procedures/settings. There is also a
factory default that is permanently stored as PST 00. see sec. 6.1

To Store the current settings press the [PST] key. The display will give the
user the option to store, recall or delete a Preset Test. The display will show
“PST STORE” or “PST RECALL” or “PST DELETE”. Use the up/down arrow
keys, [Dspl] and [Test], to select either store or recall.






If Store is selected the display will then show “PST STORE XX” with the
number showing the ID for the next empty location. For example if PST 01,
02, & 03 are already stored the next number is PST 04. If the user wishes
to overwrite a previously stored value they can change the number and
continue. Using the up/down arrow keys, [Dspl] and [Test]. If the user has
stored positions 1,2, and 4 then when this process begins the first blank
location is PST 03, so that is the value to be displayed. Then the next time
PST 05 would be shown.





Press [Setup/Enter] to select the PST number shown. If this number already
has a test method stored then ask if the test should be overwritten. To
continue the user must press [Setup/Enter] to overwrite the old results and
store new ones.


Pressing the [Tare]

key will escape without clearing the

old result.

PST STORE

PST STORE

05

OVERWRITE?