with the timer running. The timer will count up to 20:00 to allow time for the
chamber to come to a steady (room) temperature. During this time observe
the temperature in the chamber.

During this time the beeper will sound every minute to remind you that a
calibration test is in progress.

Note the temperature from the Temperature Calibration Accessory. This is
the first temperature to be entered for the lower calibration point.

When the temperature is stable and constant for some time press
[Setup/Enter] to enter the lower temperature value.

The display will change to allow the value to be entered.





Using the numeric entry method, enter the temperature inside the chamber,
for example 23.8ºC you would enter:





Press [Setup/Enter] to continue the test.

The heaters will turn on and begin heating the test chamber. The timer will
show the elapsed time as before. The chamber will be heated to
approximately 150ºC.

FIRST

00:00

FIRST

000.0

FIRST TEMP

023.8