Se [ 10, Rapid, St0p – Nova-Tech PMB Moisture Analyzer User Manual

Page 25

Advertising
background image

 

 

23

|

P a g e

© Adam Equipment Company 2009

Display “Interval” “Sec XX” The user can change the value using the arrow
keys. Range is 01 to 99 Seconds. Typical results will be computed every 5-
15 seconds.




The next setting is the operation of the Rapid heat Function.

Display will show “Rapid”




Options are to set Rapid heat to Off, On or User selectable.
The Rapid heat function is only active for single or step heating. For ramp
heating it is disabled.


The next setting is the End Point Determination or Stop Condition.
Display will show “STOP”




Options are “Time”, “Stable” or “Time/Stab”
The next steps will depend upon the option selected.

INTERVAL

se[10

OFF

rapid

ON

rapid

USER

rapid

TIME

St0p

TIME/STAB

St0p