“Ramp” Selected
Show “Ramp” the temperature as above for setting the temperature and then
change to “Ramp Time” and “ XX:XX ”, minutes and seconds.
Use the arrow keys to select the digit to change. The user can select a time
up to 99:59.



The ramp

time should be set long enough to allow the heater to warm the sample
gently.


“Step” Selected
Show “Step 1” and set the temperature of step one as for the single heat
settings above. When the value is set the display will show “Time 1” to set
the time the first step is held. Enter the value as above. The minimum time
for step 1 is 3 minutes.

When set press [Setup/Enter] to show “Step 2” and set this temperature
and time and then go to “Step 3” and set this temperature and time.
If only 2 steps are required set the time for step 3 to 00:00. The analyzer will
ignore this setting.


Interval Setting
The next setting will be to set the time interval the analyzer uses to determine
results and to print the current result. The analyzer will compute the percent
moisture or solid at the time interval desired. The new result will be displayed
and printed as described below.