10.1.3 CALIBRATION


This menu allows the user to calibrate the weight or the chamber
temperature.

Press [Setup/Enter] when “CALIBRATE” is displayed.

The options for each parameter can be scrolled through by using the [Up] or
[Down]
key

CAL MASS

Begin the mass calibration function.
Return to weighing when finished.

CAL TEMP

Begin the temperature calibration of
the weighing chamber, Return to
weighing when finished.

CAL REPORT

On = Enabled. Prints out Calibration
report after successful calibration
(either mass or temperature)
OFF = Disabled


Press [PST] to advance to setting of the next menu which is “PASSCODES”
or [Esc] to return to normal weighing.

10.1.3.1 MASS CALIBRATION
Enter the menu for Calibration section. Select CAL MASS. Press
[Setup/Enter].

Display will guide you through the steps.
Display will show “Load 0g” , Unload the pan and remove the sample pan.
Press [Setup/Enter].

The display will prompt to select the mass to be used for calibration, i.e.
“LOAD 50g”, Use the UP/DOWN arrow keys to select the alternate mass