SECTION 5: ELECTRICAL INSTALLATION
5.1. WIRING INFORMATION
•
ALL wiring cable
MUST be a minimum of
18 AWG, with individually
shielded pairs
.
─ Three (3) pair cable.
─ Use Fairbanks part number
17246
, or an equivalent.
5.1.1. Maximum Cable Lengths
No. of
LOAD CELLS
1000 Ohm Cells
Maximum Cable Lengths for
INDICATOR-TO-POWER SUPPLY
1-10
Up to 1800 feet
12
Up to 1150 feet
14
Up to 800 feet
16
Up to 575 feet
**
See
Appendix I
for the complete wiring charts of the indicator,
Power Supply (
24720
) and the Smart Sectional Controllers (
26168
).
5.2. ELECTRONIC COMPONENT CARE
•
Most of the electronic equipment consists of printed
circuit assemblies, which
must be installed using
ESD
handling procedures
.
•
All electronic and mechanical adjustments are
considered to be part of the installation, and are
included in the installation charge(s).
•
The AC receptacle/outlet must be located near the
Indicator and easily accessible.
•
Electrical connections other than those specified may
not be performed.
•
The load cells are calibrated with the cable attached,
and therefore
the cable should NOT be cut.