Help Markers Setup

1. Touch Measure → Measure Setup... on the menu bar.

2. Select a Measure Mode: Std Vertical, Std Horizontal, or My

Measure.

3. Touch the Show All button (as follows) to display Help Markers for

every parameter being measured on the displayed waveform (C2 in
the previous examples).

4. Touch a Px for any listed parameter; it doesn't matter which.

Now, on its corresponding dialog, touch inside the Help Markers
control and select Simple or Detailed. The Simple selection
produces cursors and Measure Gate gate posts. The gate posts are
independently placeable for each parameter. The Detailed selection
produces cursors, Measure Gate gate posts, a label identifying the
parameter being measured, and a level indicator and hysteresis
band for "at level" parameters.

Note: The choice of Simple or Detailed is applied to all parameters
at the same time. That is, if you choose Simple markers for one
parameter, all parameters are then displayed in this mode.

5. Touch the Always On checkbox if you want to continuously display

Help Markers for this parameter.

Turning Off Help Markers

1. Touch the Clear All button to turn off Help Markers for all

parameters.

2. Turn off Help Markers for individual parameters by touching the Px

tab for the parameter in question. Then uncheck the AlwaysOn
checkbox. When you close this dialog, the Help Markers for this
parameter are no longer shown.