WaveMaster 8 Zi/Zi-A Oscilloscopes

134

922133-00 Rev A

11. Touch the Browse button next to the Save file in directory control

and navigate to the location where you want the file saved. The file
name is assigned automatically and is shown under the control.

12. Finally, touch the Save Now! button.

Recalling Waveforms

Note: Only .trc files saved in binary format can be recalled into the
oscilloscope.

Access the Recall Waveform dialog by either selecting File → Recall
Waveform
from the menu bar or clicking the Recall Waveform button on
the main Save/Recall dialog.

Use the numbered sections of this Recall Waveform dialog, which
correspond with the following explanations, to properly recall your
waveform(s).

1. Select either Memory or File by touching either respective button on

the Recall From portion of the dialog.

Note: When Memory is selected, only Source, Destination, and
Show on recall controls are shown on the Recall Waveform dialog.
When File is selected, many more controls are available. Regardless,
the controls are used as explained in the following steps.

2. If recalling from Memory, touch inside the Source control and select

a source stored in non-volatile RAM (M1–M4) from the Select
Source
pop-up.

3. Touch inside the Destination control and select a location for

storage (memory or file) from M1 to M4.

4. Show on Recall - Mark this checkbox to have the memory trace

shown on the grid display after recalling the waveform.