Getting started manual – Teledyne LeCroy WaveMaster 8 Zi_Zi-A User Manual

Page 119

Advertising
background image

Getting Started Manual

922133-00 Rev A

111

C

OMPARING

D

UAL

P

ARAMETERS

In Dual Parameter Compare mode, your oscilloscope can take parameter
results measured on two different waveforms and compare them to each
other. If desired, set your test to be true if Any or All waveforms fit the
criteria set by the comparison condition. Setups are conveniently shown in
the Summary box of the Qx dialog as follows (shown on the Dual Param
Compare right-hand dialog):

1. Touch Analysis → Pass/Fail Setup... from the menu bar.

2. Touch one of the Qx tabs and a setup dialog for the position is

shown.

3. Touch inside the Condition field in the main dialog and select

DualParamCompare.

4. Touch inside the Source1 and Source2 fields and select a source

from the pop-up menu.

5. Touch inside the ParamCompare mini-dialog field and select a

source from the pop-up menu.

6. Touch inside the Compare Values field and select All or Any from

the pop-up menu.

By selecting All, the test are true only if every waveform falls within
the set limit. Selecting Any makes the test true if just one waveform
falls within the limit.

7. Touch inside the Condition field on the ParamCompare part of the

dialog and select a math operator from the pop-up menu.

Advertising