Aux Output

The Aux Output dialog allows you to specify Auxiliary and Calibration
Output details.

1. If you want a specialized output, touch one of the following buttons

under Use Auxiliary Output For: Square Wave, Trigger Enabled,
Trigger Out, Pass/Fail, or Off.

Note: If Pass/Fail is selected on this Use Auxiliary Output For
section of the dialog, a Pulse Duration data entry control is made
available. Provide a value within your instrument's specified range
(varied based on your specific model, refer to datasheet
specifications at teledynelecroy.com for details) using your
preferred input control method. Additional information on using the
touch screen controls can be found in the Dialog Area (on page 33).

2. On the Use Calibration Output For section of the dialog, choose

Square, DC Level, or Off. If you want a 1 kHz, 1 V Square Wave,
touch the Set to 1 kHz, 1 V Square Wave button.

Additional controls are made available based on Square or DC Level
selections as follows:

If you selected Square on the Use Calibration Output For section
of the dialog, a Frequency data entry control is made available.
You can set a value using your preferred input control method.

If you selected Square or DC Level on the Use Calibration
Output For
section of the dialog, an Amplitude into 1 MΩ data
entry control is made available. Provide a desired value using
your preferred input control method.

A

UXILIARY

O

UTPUT

S

IGNALS

A calibration signal and the following additional signals
can be output through the connector labeled AUX OUT
on your oscilloscope. Use the Auxiliary Output to send
the following signals to other instruments:

Trigger Enabled - This function can be used as a