Measurement – Teledyne LeCroy WaveMaster 8 Zi_Zi-A User Manual

Page 72

Advertising
background image

WaveMaster 8 Zi/Zi-A Oscilloscopes

64

922133-00 Rev A

G

LITCH

Glitch trigger is a simpler form of Width trigger. Use Glitch trigger when
you want to define a fixed pulse-width time or time range only. Glitch
trigger makes no provision for voltage levels or ranges.

D

ROPOUT

Used primarily in single-shot applications, and usually with a pre-trigger
delay, Dropout trigger can detect lost signals. The trigger is generated at
the end of the timeout period following the last trigger source transition.

R

UNT

The Runt trigger occurs when a pulse crosses a first threshold line, but fails
to cross a second threshold line before re-crossing the first. Other defining
conditions for this trigger are the edge (triggers on the slope opposite to
that selected) and runt width.

S

LEW

R

ATE

SlewRate trigger activates a trigger when the rising or falling edge of a
pulse crosses two threshold levels: an upper level and a lower level. The
pulse edge must cross the thresholds faster or slower than a selected
period of time.

Measurement

This trigger type allows you to leverage parameter measurements as
waveform trigger conditions. A measurement trigger is either the only
trigger or the final trigger in a chain of trigger events including hardware
triggers.

Advertising