WaveMaster 8 Zi/Zi-A Oscilloscopes

74

922133-00 Rev A

Training TriggerScan

The TriggerScan Trainer inspects a currently acquired waveform and
automatically builds a list of common trigger setups used to find rare
events.

PLEASE NOTE THE FOLLOWING:

You must acquire and display at least 3 cycles of a signal before
running the Trainer.

You should run the Trainer if you want to change the trigger types or
if you change the channel or signal.

1. Touch Trigger → Trigger Setup... from the menu bar.

2. On the Trigger dialog, click the TriggerScan tab.

3. Touch the Trainer button and select a channel as the source for the

training on the Trigger Scan Trainer pop-up. Also choose the types
of triggers the Trainer should use to train the system and then touch
the Start Training button. The training begins. When it is complete, a
list of smart trigger setups is displayed in the Trigger List.