Environmental conditions, Probing characteristics, Switches – Teledyne LeCroy USB Protocol Suite User Manual (Voyager_Advisor T3_Mercury) User Manual
Page 52: Recording memory size

Teledyne LeCroy Corporation
Voyager M3x Analyzer
52
USB Protocol Suite User Manual
Environmental Conditions
Probing Characteristics
Switches
Recording Memory Size
1 GB or 4 GB for traffic data capture, timing, state and other data.
Operating
Temperature
0 to 55 °C (32 to 131 °F)
Storage Range
‐20 to 80 °C (‐4 to 176 °F)
Operating Humidity
10 to 90%, non‐condensing
Operating Altitude
Up to 6560 feet (2000 meters)
Connection
SuperSpeed connectors: The Voyager SuperSpeed
(3.0) analyzer and exerciser port differential input
impedance is between 80 and 120 ohms.
USB 3.0 cables
High Speed USB Connectors
Standard cables
Note: The USB 2.0 exerciser port is not terminated
differentially. The single ended termination is
45 ohms to ground +/‐ 10% or between 40.5
and 49.5 ohms. The effective differential USB
2.0 termination is between 80 and 100 ohms
(90 ohms +/‐ 10%).
Power
On/off