Warning, 4 iecex certification, 4) operation – Yokogawa EJX115A User Manual

Page 16: 5) maintenance and repair, 6) name plate, A. iecex intrinsically safe type / type n, Handling cautions, 9-32 nakacho, musashino-shi, tokyo japan

Advertising
background image

<2. Handling Cautions>

14

IM 01C25A01-01E

(4) Operation

WARNING

• OPEN CIRCUIT BEFORE REMOVING COVER.

INSTALL IN ACCORDANCE WITH THIS USER’S
MANUAL

• Take care not to generate mechanical sparking

when access to the instrument and peripheral
devices in a hazardous location.

(5) Maintenance and Repair

WARNING

The instrument modification or parts replacement by
other than an authorized Representative of Yokogawa
Electric Corporation is prohibited and will void the
certification.

(6) Name Plate

F0215.ai

Name plate

Tag plate for flameproof type

Tag plate for intrinsically safe Ex ia

WARNING

D

D

WARNING

No. KEMA 07ATEX0109 X

Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db

Enlcosure : IP66/IP67

TEMP. CLASS T6 T5 T4

MAX PROCESS TEMP.(Tp.) 85 100 120 °C

Tamb. -50 to 75 80 75 °C

T85°C(Tamb.:-30(-15) to 75°C, Tp.:85°C)(for Dust)

No. DEKRA 11ATEX 0228 X

Ex ia IIC T4 Ga Ta: -50 TO 60°C

Ex ia IIIC T85°C T100°C T120°C Db Ta:-30(-15) TO 60°C

IP66/IP67

MAX. PROCESS TEMP.(Tp.) 120°C

T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)

Ui=30V, Ii=200mA , Pi=0.9W, Ci=27.6nF, Li=0µH

POTENTIAL ELECTROSTATIC

CHARGING HAZARD

- SEE USER’S MANUAL

AFTER DE-ENERGIZING, DELAY 5 MINUTES BEFORE

OPENING.

WHEN THE AMBIENT TEMP.

65°C, USE THE

HEAT-RESISTING CABLE & CABLE GLAND

≥ 90°C

POTENTIAL ELECTROSTATIC CHARGING HAZARD

Tag plate for intrinsically safe type

(for EJX910A and EJX930A)

D

No. KEMA 06ATEX0037 X

EEx ia IIC T4

IP66 and IP67

Tamb. -50(-15) to 60°C MIN Tamb.for DUST -40(-15°C)

MAX PROCESS TEMP.(Tp) 120°C

T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)

Supply/Pulse circuit

Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH

Sensor circuit

Uo=30 V, Io=95.4 mA, Po=468 mW, Co=11 nF, Lo=3.9 mH

Tag plate for intrinsically safe Ex ic

WARNING

Ex ic IIC T4 Gc

IP66

Tamb -30(-15) TO 60°C

MAX. PROCESS TEMP. 120°C

Ui=30V, Ci=27.6nF, Li=0µH

POTENTIAL ELECTROSTATIC

CHARGING HAZARD

- SEE USER’S MANUAL

*3

*3

*3

F0215.ai

Name plate

Tag plate for flameproof type

Tag plate for intrinsically safe Ex ia

WARNING

D

D

WARNING

No. KEMA 07ATEX0109 X

Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db

Enlcosure : IP66/IP67

TEMP. CLASS T6 T5 T4

MAX PROCESS TEMP.(Tp.) 85 100 120 °C

Tamb. -50 to 75 80 75 °C

T85°C(Tamb.:-30(-15) to 75°C, Tp.:85°C)(for Dust)

No. DEKRA 11ATEX 0228 X

Ex ia IIC T4 Ga Ta: -50 TO 60°C

Ex ia IIIC T85°C T100°C T120°C Db Ta:-30(-15) TO 60°C

IP66/IP67

MAX. PROCESS TEMP.(Tp.) 120°C

T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)

Ui=30V, Ii=200mA , Pi=0.9W, Ci=27.6nF, Li=0µH

POTENTIAL ELECTROSTATIC

CHARGING HAZARD

- SEE USER’S MANUAL

AFTER DE-ENERGIZING, DELAY 5 MINUTES BEFORE

OPENING.

WHEN THE AMBIENT TEMP.

65°C, USE THE

HEAT-RESISTING CABLE & CABLE GLAND

≥ 90°C

POTENTIAL ELECTROSTATIC CHARGING HAZARD

Tag plate for intrinsically safe type

(for EJX910A and EJX930A)

D

No. KEMA 06ATEX0037 X

EEx ia IIC T4

IP66 and IP67

Tamb. -50(-15) to 60°C MIN Tamb.for DUST -40(-15°C)

MAX PROCESS TEMP.(Tp) 120°C

T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)

Supply/Pulse circuit

Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH

Sensor circuit

Uo=30 V, Io=95.4 mA, Po=468 mW, Co=11 nF, Lo=3.9 mH

Tag plate for intrinsically safe Ex ic

WARNING

Ex ic IIC T4 Gc

IP66

Tamb -30(-15) TO 60°C

MAX. PROCESS TEMP. 120°C

Ui=30V, Ci=27.6nF, Li=0µH

POTENTIAL ELECTROSTATIC

CHARGING HAZARD

- SEE USER’S MANUAL

*3

*3

*3

MODEL: Specified model code.

STYLE: Style code.

SUFFIX: Specified suffix code.

SUPPLY: Supply voltage.

OUTPUT: Output signal.

MWP: Maximum working pressure.

CAL RNG: Specified calibration range.

NO.: Serial number and year of production

*1

.

TOKYO 180-8750 JAPAN:

The manufacturer name and the address

*2

.

*1: The first number in the second block of “NO.” column

is the last one number of the production year.

second block

NO. 91K819857 132 7

The year 2011

*2: “180-8750” is a zip code which represents the

following address.

2-9-32 Nakacho, Musashino-shi, Tokyo Japan

*3: The identification number of Notified Body.

2.4.4 IECEx Certification

Model EJX Series pressure transmitters with

optional code /SU2 can be selected the type of

protection (IECEx Intrinsically Safe/type n or

flameproof) for use in hazardous locations.

Note 1. For the installation of this transmitter, once a

particular type of protection is selected, any

other type of protection cannot be used. The

installation must be in accordance with the

description about the type of protection in this

instruction manual.

Note 2. In order to avoid confusion, unnecessary

marking is crossed out on the label other

than the selected type of protection when the

transmitter is installed.

a. IECEx Intrinsically Safe Type / type n

Caution for IECEx Intrinsically safe and type n.

Note 1. Model EJX Series differential, gauge, and

absolute pressure transmitters with optional

code /SU2 are applicable for use in hazardous

locations

• No. IECEx CSA 05.0005

• Applicable Standard: IEC 60079-0:2000,

IEC 60079-11:1999, IEC 60079-15:2001

Advertising