7 low voltage directive, Low voltage directive – Yokogawa EJX115A User Manual

Page 19

Advertising
background image

<2. Handling Cautions>

17

IM 01C25A01-01E

2.7 Low Voltage Directive

Applicable standard: EN61010-1

(1) Pollution Degree 2

"Pollution degree" describes the degree to which

a solid, liquid, or gas which deteriorates dielectric
strength or surface resistivity is adhering. " 2 " applies
to normal indoor atmosphere. Normally, only non-
conductive pollution occurs. Occasionally, however,
temporary conductivity caused by condensation must
be expected.

(2) Installation Category I

"Overvoltage category(Installation category)"

describes a number which defines a transient
overvoltage condition. It implies the regulation for
impulse withstand voltage. " I " applies to electrical
equipment which is supplied from the circuit when
appropriate transient overvoltage control means
(interfaces) are provided.

H

an

dlin

g C

au

tio

ns

2

Advertising