4 mounting the flushing connection ring, Mounting the flushing connection ring, Important – Yokogawa EJX115A User Manual

Page 22

Advertising
background image

<3. Installation>

20

IM 01C25A01-01E

IMPORTANT

The transmitter should be installed at least 600 mm below
the high pressure (HP) process connection to ensure a
positive head pressure of fill fluid. Pay special attention to
vacuum applications.
If it can not be installed at least 600 mm below the HP
process connection, please use the equation below:

h=

× 0.102

[mm]

(P–P0)

ds

h: Vertical height between the HP process connection

and the transmitter (mm)
h≤0: Install the transmitter at least h (mm) below the

HP process connection

h>0: Install the transmitter at most h (mm) above the

HP process connection

P: Pressure in the tank (Pa abs)
P0: Minimum working pressure limit of the transmitter (Pa

abs) See below table.

[For fill fluid code A, B, C, D, E]

Wetted parts

material

code

Capillary

length

Process connection size

code

2, 8

3

4,W

SW

1 to 5m

6790

3190

6 to 10m

10030

3520

SE

1 to 5m

6790

3190

6 to 10m

10030

3520

SY

1 to 5m

3190

6 to 10m

3520

HW

1 to 5m

19150

6140

6 to 10m

8290

TW

1 to 5m

9620

3620

6 to 10m

4210

UW

1 to 5m

9540

4750

6 to 10m

6050

[For fill fluid code 1, 2, 4]

Wetted parts

material

code

Capillary

length

Process connection size

code

2, 8

3

4,W

SW

1 to 5m

2570

320

6 to 10m

4680

530

SE

1 to 5m

2570

320

6 to 10m

4680

530

SY

1 to 5m

320

6 to 10m

530

HW

1 to 5m

10220

2050

6 to 10m

3450

TW

1 to 5m

4270

570

6 to 10m

960

ds: Specific gravity of fill fluid (at 25°C). See below table.

Fill fluid

code

A, 1, 4

B

C, 2

D

E

ds: Specific

gravity

1.07

0.94

1.09

1.90 to

1.92

1.09

F0308.ai

P

Low pressure side

High

pressure

side

(+)

(–)

0

h

Figure 3.8

Example of Installation to Tank

(Caution on Installation)

3.4 Mounting the Flushing

Connection Ring

3.4.1 Mounting to Pressure Detector

Section

The flushing connection ring is mounted to the pressure
detector section as shown in Figure 3.9.
At the factory shipment, the flushing connection ring is
already assembled and attached to process detector
section.

Ring holder

Spiral gasket

F0309.ai

View from pressure

detector section

Groove for

installing

spiral gasket

Pressure-detector section

Drain/vent plug

Ring

Figure 3.9

Mounting to Pressure Detector Section

(1) Mount the ring holder on the ring and loosely tighten

the mounting screws.

(2) Place the spiral gasket in the ring groove. With the

ring correctly aligned and flush with the face of the
pressure detector, securely tighten each ring holder’s
mounting screws.

(3) Position the ring so that the drain/vent plugs are

aligned straight up and down.