3 wiring, Wiring, 1 loop configuration – Yokogawa EJX115A User Manual

Page 32: 2) intrinsically safe type, Simultaneous analog-pulse output, Figure 5.5 external temperature connection

Advertising
background image

<5. Wiring>

30

IM 01C25A01-01E

5.2.5 External Temperature Connection

(for EJX910A and EJX930A)

Connect the RTD cable assembly to the Juck Terminal.

F0505.ai

SUPPL

Y

PULSE

CHECK

ALARM

Figure 5.5

External Temperature Connection

5.3 Wiring

5.3.1 Loop Configuration

Since the DPharp uses a two-wire transmission system,
signal wiring is also used as power wiring.
DC power is required for the transmitter loop. The
transmitter and distributor are connected as shown below.
For details of the power supply voltage and load
resistance, see Section 5.6.

(1) General-use Type and Flameproof Type

• Analog Output

Hazardous Location

Nonhazardous Location

Transmitter terminal box

Distributor

(Power supply unit)

Receiver

instrument

F0506.ai

SUPPL

Y

PULSE

CHECK

ALARM

Figure 5.6

Connection between Transmitter and

Distributor

• Pulse output and Alarm, Status Output or

Simultaneous Analog-Pulse Output

(for EJX910A and EJX930A)

SUPPL

Y

PULSE

CHECK

ALARM

Hazardous Location

Nonhazardous Location

Transmitter terminal box

Distributor

(Power supply unit)

Receiver

instrument

Receiver

instrument

F0507.ai

Figure 5.7

Connection between Transmitter and

Distributor

(2) Intrinsically Safe Type

With the intrinsically safe type, a safety barrier must be
included in the loop.

Hazardous Location

Nonhazardous Location

Transmitter terminal box

Distributor

(Power supply unit)

Receiver

instrument

Safety barrier

F0508.ai

SUPPL

Y

PULSE

CHECK

ALARM

Figure 5.8

Connection between Transmitter and

Distributor

(3) Intrinsically Safe Type (for EJX910A and

EJX930A)

• Analog Output

SUPPL

Y

PULSE

CHECK

ALARM

Hazardous Location

Nonhazardous Location

Transmitter terminal box

Receiver instrument

(and power supply)

Safety barreir

(Isolation type)

F0509.ai

Anaolg input

Power

supply

• Simultaneous Analog-Pulse Output

SUPPL

Y

PULSE

CHECK

ALARM

Hazardous Location

Nonhazardous Location

Transmitter terminal box

Safety barrier

(Isolated type)

I/O module

F0510.ai

Safety barrier

(Isolated type)

Power

supply

Power

supply

Analog input

Pulse input

Figure 5.9

Connection between Transmitter,

barrier and receiver

Advertising