Operation, 1 preparation for starting operation, Preparation for starting operation – Yokogawa EJX115A User Manual

Page 37

Advertising
background image

<6. Operation>

35

IM 01C25A01-01E

6. Operation

NOTE

For F

OUNDATION

Fieldbus and PROFIBUS PA

communication types and for the transmitter operating
confirmation and zeroing by any communication
method, refer to manuals in the attached CD-ROM for
further information.

6.1 Preparation for Starting

Operation

Confirming that Transmitter is Operating

Properly

On the integral indicator

• If the wiring system is faulty, the display stays blank.
• If the transmitter is faulty, an error code is displayed.

Self-diagnostic error on the integral indicator

(Faulty transmitter)

F0601.ai

NOTE

If any of the above errors are indicated on the display
of the integral indicator, refer to Chapter 7 for the
corrective action.

Verify and Change Transmitter

Parameter Setting and Values

The parameters related to the following items are set at
factory as specified in order.

• Calibration range
• Integral indicator display
• Output mode
• Software damping (optional)

Other parameters like following are shipped with the
default setting.

• Low-cut
• Process alarm setting
• Static pressure range
• Signal characterizer
• Write protection

To confirm or change the values, please refer to manuals
in the attached CD-ROM.

Output Status Setting at CPU Failure and

Hardware Write Protection

Set the switches as shown in the figure below to set the
burn-out direction and write protection. The Burnout
switch is set to the H side for delivery (unless option code
/C1 or /C2 is specified in the order), and the hardware
write protection switch is set to E side. The setting of the
switches can be confirmed via communication.

F0602.ai

H

L

H

L

Slide switch

CPU assembly

BO H

L

WR E

D

Burnout direction switch

Write protection switch

H

L

H

L

E

D

E

D

Hardware write protection switch (WR)

Write Protection

Switch Position

HIGH

LOW

Burnout direction switch (BO)

Burnout Direction

Burnout Direction

Switch Position

YES

(Write disabled)

NO

(Write enabled)

Write Protection

Figure 6.1

Burn-out Direction and Hardware Write

Protection Slide Switch

O

pe

ra

tio

n

6

Advertising