App index – Yokogawa Removable Chassis DX1000N User Manual

Page 121

Advertising
background image

Measurement Channels and

Alarms

IM 04L41B01-01E

3-27

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

3.12 Using the Alarm Annunciator Function (Release number 3 or later)

ISA-A

Sequence diagram

Process

Sequence

Display

Relay output

Alarm icon

Normal
Normal
Normal

Deactivated

No display

Process

Sequence

Display

Relay output

Alarm icon

Deactivated

Process

Sequence

Display

Relay output

Alarm icon

Activated

Abnormal

Acknowledged

On

On (red)

Blinking (red)

Flashing

Alarm

Abnormal or

normal

To abnormal

Return to normal

Alarm ACK

while abnormal

Alarm ACK

while normal

Sequence pattern

Alarm

Display

Relay output

Alarm ACK

Alarm ACK

If an alarm is released before the alarm ACK operation is performed, the annunciator blinks with the color that is displayed

when no alarms are activated.

If an alarm is released before the alarm ACK operation is performed, the alarm icon blinks green.

When multiple alarms have been assigned to a single annunciator window

If multiple alarms are occurring, the window is displayed using the color of the alarm with the highest priority.

When two alarms have been assigned to a single relay

When the relay is set to OR logic

ACK

ACK

Relay hold

Alarm A

Occurrence

Alarm B

Occurrence

Release

Release

Relay

Activated

Deactivated

An alarm has occurred so the relay has been activated.