App index action – Yokogawa Removable Chassis DX1000N User Manual

Page 50

Advertising
background image

1-39

IM 04L41B01-01E

Overview of Functions

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

Action

• Actions

Select from the following actions.

Action

Level/Edge

*1

Description

Memory start/stop

Level

Starts/stops memory sampling.

Memory start

Edge

Starts the memory sampling.

Memory stop

Edge

Stops the memory sampling.

Event trigger

*4

Edge

Applies a trigger for starting the event data

recording.

This is valid when recording event data in trigger

mode. See the next page.

Alarm acknowledge

Edge

Releases the alarm output.

This is valid when the use of the alarm ACK

operation is enabled.

Computation start/stop

*2

Level

It is used to start/stop the computation.

Computation start

*2

Edge

Starts the computation.

Computation stop

*2

Edge

Stops the computation.

Computation reset

*2

Edge

Resets the computed values on all computation

channels.

Save display data

Edge

The display data being recorded is saved to the

internal memory as a file. This is the same function

as the data save operation using the FUNC key.

Save event data

Edge

The event data being recorded is saved to the

internal memory as a file.

This is the same function as the data save operation

using the FUNC key.

Message

Edge

Writes a message.

This action can be executed while memory sampling

is in progress.

Snapshot

Edge

Saves the screen image data.

Switch the display rate

Level

Toggles between the trend interval and the

secondary trend interval.

This action is valid when the DX is configured to use

trend interval switching.

Manual sample

Edge

Executes manual sampling.

Reset the relative timer

Edge

Resets the relative timer. The timer restarts from

that point.

Switch the display group

Edge

Switches the display group when the trend, digital,

or bar graph is displayed.

Flag

*2

Level

The flag is zero for normal conditions and 1 when an

event occurs. When the event is an edge operation,

the value alternates between 0 and 1 whenever an

event occurs.

The flag can be written in a calculation expression

of a computation channel.

Load the setup

*4

Edge

Loads the setup data file in the root directory of the

CF card into the DX and updates the DX settings.

See below.

Adjust the time

Edge

Synchronizes the time to the nearest hour. See the

next page.

Display comment

*3

Edge

Displays a preset text string (comment text block).

For information about how to set comment text

blocks, see section 5.18.

Show favorite display

*3

Edge

Shows the display that is registered to the Favorite

key. For information about how to register a display

to the Favorite key, see section 5.15.

Reset alarm display

*3

Edge

An operation for when you are using a double lock-

in sequence (ISA-M) with the alarm annunciator

function. Resets the alarm display. For information

about annunciator settings, see section 3.12.

*1 For a description of level and edge, see “Miscellaneous” in this section.
*2 This is an option.
*3 This function is available for release numbers 3 and later.
*4 This action is not available on models with the /AS1 advanced security option.

1.6 Event Action and Remote Control Functions (/R1 and /PM1 Options)