Yokogawa Removable Chassis DX1000N User Manual

Page 270

Advertising
background image

10-2

IM 04L41B01-01E

Code

Message

Explanation/Countermeasures/Ref. section

43

A channel number cannot repeat in a group.

Check that a channel is not registered twice.
See section 5.1.

45

There is no character string saved in the clipboard.

Copy a character string to the clipboard.

46

The character string saved in the clipboard is too long.

Paste a character string with the specified number of

characters.

47

Start and end time cannot match.

Check the starting and ending times.
See section 2.1.

48

Invalid or missing DST time settings.

Check the starting and ending times.
See section 2.1.

61

There is no channel specified by the MATH expression.

Check the channel number specified by the expression.

See sections 1.8 and 9.1.

62

MATH expression grammar is incorrect.

Check that the expression grammar is correct.
See section 9.2.

63

MATH expression sequence is incorrect.

Check that the operator used in the expression in

relation to the applicable operands meets the grammar

requirements.
See section 9.2.

64

MATH upper and lower span values are equal.

Set the upper limit not equal to the lower limit.
See section 9.1.

65

Too many operators for MATH expression.

The maximum number of operators in an expression

has been exceeded. Reduce the number of operators,

such as by splitting up the expression into multiple

computation channels.See section 9.2.

70

Nonexistent constant specified in MATH expression.

Check the constant number specified by the expression.
See section 9.1.

71

Set range of the MATH constant is exceeded.

Check the selectable range.
See section 9.1.

80

This username is already registered.

Register another user name.
See section 8.2.

81

All space or ‘quit’ string cannot be specified.

Change the character string.
See section 8.2.

83

This user ID and password combination is already in use.

Change the user ID or the password.
See section 2.1 in the advanced security manual.

84

The login password has not been set up.

Set a password.
See section 8.2.

85

The login password is incorrect.

Check the password. If you lost the password, ask your

administrator to reset it.
See sections 8.2 and 8.3.

86

The key-lock release password is incorrect.

Check the password. If you lost the password, it must

be reset.
See section 8.1.

87

This key is locked.

Release the key lock.
See section 8.1.

88

This function is locked.

Release the key lock.
See section 8.1.

89

Press [FUNC] key to login.

Log in.
See section 8.3.

90

No permission to enter to the SETUP mode.

Check the keylock or login settings.
See sections 8.2 and 8.3.

91

Password is incorrect.

Enter the correct password. If you lost the password, it

must be reset.
See sections 8.2 and 8.3.

92

Press [ESC] key to change to the operation mode.

Press the ESC key.

93

String including space or all space cannot be specified.

Spaces are not allowed in the Web browser user name

and password.
See section 1.5 in the communication manual.

94

More than one address cannot be specified.

Only a single sender is allowed.
See section 1.4 in the communication manual.

95

Number entered exceeds channel number range.
Use another command.

Check the syntax of the Modbus command.
See sections 1.10 and 2.6 in the communication

manual.

10.1 A List of Messages

Advertising
This manual is related to the following products: