Yokogawa Removable Chassis DX1000N User Manual

Page 124

Advertising
background image

3-30

IM 04L41B01-01E

3.13 Managing the Input Calibration Interval (/CC1

option; release numbers 4 and later)

Manage the input calibration interval to make sure that calibration is performed regularly.
For a description of the function, see section 1.1.

Setup Screen

Press

MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic

setting mode), and select the

Menu tab > Set Calibration.

Setup Items

• Set Calibration > Use/Not

Select whether to

Use or Not use calibration management.

When you select Use, the Set Calibration tab appears in setting mode.

• Set Calibration > Notification (DayBfr)

You can specify how many days before the calibration due date you want to display

the calibration notification screen. You can set the number of days to a value between
1 and 10.

• Set Calibration > Renotification

You can specify the period at which to display the calibration notification screen. The

calibration notification screen will continue to appear until calibration is completed.

Setting

Description

10min

The screen will appear every 10 minutes.

30min

The screen will appear every 30 minutes.

1h

The screen will appear every hour.

8h

The screen will appear every 8 hours.

24h

The screen will appear every 24 hours.