Yokogawa Removable Chassis DX1000N User Manual

Page 247

Advertising
background image

9-3

IM 04L41B01-01E

Computation and Report Functions (/M1 and /PM1 Options)

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

• Span_L, Span_U

Set the measurement range.
Selectable range of values: –9999999 to 99999999
Selectable decimal places: X.XXXX, XX.XXX, XXX.XX, XXXX.X

• Unit

Set the unit of the computed value (Up to 6 characters,

A a # 1

).

• Alarm

The available alarm types are high limit alarm, low limit alarm, delay high limit alarm,

and delay low limit alarm.

The range of alarm values is as follows:

Type

Value

H, L, T, t

Within –9999999 to 99999999 excluding the decimal point

For details on setting alarms, see section 3.7.

* If the Math On/Off or calculation expression is changed, the alarms for that channel are turned

Off.

• Alarm delay > Time

Set the alarm delay time using an integer in the range of 1 to 3600 s.

• Tag > Comment

Sets the tag comment. On a DX with a release number of 3 or later, you can enter up

to 32 characters. On a DX with a release number of 2 or earlier, you can enter up to
16 characters. The characters that can be entered are:

A a # 1

.

• Tag > No. (Release number 3 or later)

This setting only appears when you have enabled the use of tag numbers.
Sets the tag number. (Up to 16 characters:

A a # 1

)

• Constant

• Number of constant

Select the constant (K01 to K60) to set.

• Value

The selectable range is as follows:
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
The number of significant digits of a constant is five. When specifying the constant

using exponential notation, set the mantissa less than or equal to 5 digits and the
exponent less than or equal to 2 digits.

• TLOG

• Timer type

You can set the timer type to

Timer or Match T (match time timer).

• Timer No.

Select the timer number to use.
For details on setting the timer, see section 7.1.
For match time timer setting instructions, see section 7.1.

• Sum scale

Set the sum scale to

/s to /h to match the unit of the measured value.

Example: If the unit of the measured value is “m

3

/min,” select

/min.

Off:

Sums as-is the measured data per scan interval.

• Reset

To reset the TLOG computed value at each interval, select

On.

9.1 Setting the Expression, Measurement Range, Alarm, Tag, and Data Storage on Computation Channels

Advertising
This manual is related to the following products: