Yokogawa Removable Chassis DX1000N User Manual

Page 259

Advertising
background image

9-15

IM 04L41B01-01E

Computation and Report Functions (/M1 and /PM1 Options)

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

• Stopping the Computation

• Stopping the Computation Simultaneously with the Memory Sampling

1.

Press

STOP.

A confirmation dialog box appears.

2.

Select

Mem+Math and press DISP/ENTER.

The memory sampling and computation stop, and the computation icon in the status display
section disappears.

• Stopping Only the Computation

1.

In the operation mode, press

FUNC.

The FUNC key menu appears.

2.

Press the

Math stop soft key.

The computation stops, and the computation icon in the status display section disappears.

Note

When the computation is stopped, the computed data of the computation channel is held at the
value that existed immediately before the computation is stopped. When memory sampling is in
progress, the held value is recorded.

• Resetting the Computed Results on All Computation Channels

This operation can be carried out when the computation is stopped. You can carry out this

operation even while the computation is in progress on DXs with release number 2 or later.

1.

In the operation mode, press

FUNC.

The FUNC key menu appears.

2.

Press the

Math reset soft key.

The computed results of all computation channels are reset.

• Releasing the Computation Data Dropout Display

This operation can be carried out when a computation data dropout occurs. When a

computation data dropout occurs, the computation icon turns yellow.

1.

In the operation mode, press

FUNC.

The FUNC key menu appears.

2.

Press the

Math ACK soft key.

The computation icon returns to white.

*

Math ACK is displayed in the FUNC key menu only when a computation data dropout
occurs.

Note

A computation data dropout occurs when the computation process cannot be completed within
the scan interval. If computation data dropout occurs frequently, lessen the load on the CPU by
reducing the number of computation channels or setting a longer scan interval. If a computation
data dropout occurs during memory sampling, the data immediately before the dropout is
recorded as the computed data of the scan interval in which the dropout occurred.

9.4 Starting/Stopping Computation, Resetting Computation, and Releasing Computation Data Dropout

Display

Advertising
This manual is related to the following products: